Share Email Print
cover

Proceedings Paper

Impedance-controlled ultrasound probe
Author(s): Matthew W. Gilbertson; Brian W. Anthony
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

An actuated hand-held impedance-controlled ultrasound probe has been developed. The controller maintains a prescribed contact state (force and velocity) between the probe and a patient's body. The device will enhance the diagnostic capability of free-hand elastography and swept-force compound imaging, and also make it easier for a technician to acquire repeatable (i.e. directly comparable) images over time. The mechanical system consists of an ultrasound probe, ball-screw-driven linear actuator, and a force/torque sensor. The feedback controller commands the motor to rotate the ball-screw to translate the ultrasound probe in order to maintain a desired contact force. It was found that users of the device, with the control system engaged, maintain a constant contact force with 15 times less variation than without the controller engaged. The system was used to determine the elastic properties of soft tissue.

Paper Details

Date Published: 25 March 2011
PDF: 12 pages
Proc. SPIE 7968, Medical Imaging 2011: Ultrasonic Imaging, Tomography, and Therapy, 796816 (25 March 2011); doi: 10.1117/12.882169
Show Author Affiliations
Matthew W. Gilbertson, Massachusetts Institute of Technology (United States)
Brian W. Anthony, Massachusetts Institute of Technology (United States)


Published in SPIE Proceedings Vol. 7968:
Medical Imaging 2011: Ultrasonic Imaging, Tomography, and Therapy
Jan D'hooge; Marvin M. Doyley, Editor(s)

© SPIE. Terms of Use
Back to Top